1
Installation
Column Installation Requirements
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
53
To install the adapters for encapsulated graphite ferrules
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
4. Remove the analytical column.
a.
Open the front door of the GC.
b. Loosen the retaining nut from the injector and detector fitting on the upper interior
wall of the GC oven.
c.
Remove the analytical column with its nut and ferrule from the bottom of the
injector and the detector.
5. Remove the bottom parts of the SSL injector. See
.
Table 9.
Adapters and 1/2-in. Wrench
Part
Description
Adapter for connection to the SSL injector
Adapter for connection to the FID, NPD, ECD, TCD, or
FPD detector
1/2-in. wrench
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.