Installation
Installing the Autosampler
38
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
To connect a TriPlus RSH/TriPlus 100 Liquid Sampler autosampler
1. Use the Ethernet cable to connect the LAN connector on the back of the TriPlus
RSH/TriPlus 100 Liquid Sampler Control Board to the computer, and verify that the
LAN control LED lights up when connected to a powered-on computer.
2. Using the Y-shape handshake cable provided, connect:
a.
the main connector to the connection marked
INTERFACE
on the back of the
control interface of the TriPlus RSH/TriPlus 100 Liquid Sampler.
b. the first branch of the Y-shape handshake cable, labelled
GC 1
to the marked
AUTOSAMPLER HANDSHAKE
on the back panel of the GC.
3. Set the Primary/Secondary switch on
PRIMARY
position.
4. Establish communication between the TriPlus RSH/TriPlus 100 Liquid Sampler and the
GC following the instructions detailed in the
TriPlus RSH Hardware Manual
or
TriPlus
100 Liquid Sampler Hardware Manual
.
To connect a second GC coupled with the same TriPlus RSH/TriPlus 100 Liquid
Sampler autosampler
1. Connect the second branch of the Y-shape handshake cable, labelled
GC 2
,
to the
connector marked
AUTOSAMPLER HANDSHAKE
on the back panel of the second
GC.
2. Set the Primary/Secondary switch on
SECONDARY
position.
3. Establish communication between the TriPlus RSH/TriPlus 100 Liquid Sampler and the
GC following the instructions detailed in the
TriPlus RSH Hardware Manual
or
TriPlus
100 Liquid Sampler Hardware Manual
.
To connect a a TriPlus autosampler
1. Using the cable provided, connect the 6-pin female connector marked
HANDSHAKE
MAIN
located on the back portion of the crossrail X and the connector marked
AUTOSAMPLER HANDSHAKE
on the GC back panel.
2. Using the cable provided, connect the 9-pin female connector marked
RS232-1
located
on the back portion of the crossrail X to 9-pin connector marked
AUTOSAMPLER 1
on
the GC back panel.
3. Connect the power cable to the AC Input connector (Mains socket) on the autosampler
and to the wall outlet.
To connect a second GC is also coupled with the same TriPlus autosampler
1. Using the cable provided, connect the 6-pin female connector marked
HANDSHAKE
VIRTUAL
located on the back portion of the crossrail X to the connector marked
AUTOSAMPLER HANDSHAKE
on the back panel of the second GC.