Detectors Advanced Maintenance
Cleaning or Replacing the FID Jet
274
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
b. Loosen the polarizing electrode fixing screw using a T10 Torxhead screwdriver, and
extract the polarizing electrode from its seat. See
.
Figure 252.
FID Jet (2)
c.
Place the polarizing electrode on a clean surface.
10. Remove, clean, and reinstall the jet.
a.
Use the elbowed box wrench for unscrewing and extracting the jet through the
detector body. See
.
Figure 253.
FID Jet (2)
WARNING
The collector insulator must be stored somewhere it can stay clean.
Jet
Polarizing Electrode
Fixing Screw
Polarizing Electrode
Jet