1
Installation
Making the Gas Supply Plumbing Connections
8
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
To plumb the gas supply
1. By now you must have done the following:
a.
Completely built your gas supply lines including any traps, tees, and extra tubing to
allow about 40 cm (16 in.) of slack in the line.
b. Purged the gas line after every tube cut to remove any debris or contaminants.
c.
Ensured the gas supply is turned off.
2. Connect the gas lines.
a.
TRACE 1300/TRACE 1310 is provided with six gas inlet ports for the connection of
carrier and detector gases. See
.
Figure 4.
Gases Inlet Connections
Figure 5.
Fittings for Gas Inlets Connection
Note
Use the 1/8-in. Swagelok fittings provided on the gas inlet ports (see
to connect the gas lines.
Gas
In
let
Ports
Carrier Front
Hydrogen
Air
Make-up Back
De
te
ct
or
G
ase
s
Carrier Back
Make-up Front
Gas Inlet Fitting
Ferrules
Nut
O-Ring