1
Installation
Column Installation Requirements
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
55
Figure 63.
Detector Bottom Parts Removal
a.
Using the 1/2-in. wrench provided, unscrew the retaining nut from the bottom of the
detector.
8. Install the adapter on the bottom the detector. See
.
Figure 64.
Adapter Installation on the Bottom of the Detector
a.
Fingertighten the retaining nut of the adapter until it starts to grip the bottom of the
detector.
b. Use the 1/2-in. wrench to firmly tighten the retaining nut of the adapter with the
base seal and the washer from the bottom of the injector.
9. Connect the column.
a.
Before starting, use a 1/4-5/16-in. wrench to remove the column nuts and ferrules
from the base of the injector and detector adapters. See
.
Note
Save the bottom parts of the detector in a safe place because they will be reused
when you restore the original configuration.
Detector Bottom
Retaining Nut
FID Graphpack
Adapter
Silver Gasket