9
Adding Modules
Adding a GSV Injector Module
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
425
19. If present, move the autosampler towards the module to restore the original alignment.
Adding a GSV Injector Module
This section provides instructions for adding a front/back Gas Sampling Valve injector
module. According to the configuration of your TRACE 1300/TRACE 1310, a dummy
module is present into the free site where the injector module is not installed.
See
Figure 440.
Add a New Front/Back GSV Injector Module
To connect the gas sampling valve module to the TRACE 1300/1310 GC
1. Put the GC in standby condition.
2. Cool the oven, injector, and detector to room temperature.
3. Close the gas supplies.
4. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
Dummy Front Injector Module
Dummy Back Injector Module
ATTENTION
Where a dummy module is installed, the gas connection is blocked by a
plug.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.