8
Installing Optional Kits
Installing the Merlin Microseal High Pressure Valve Kit
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
331
Figure 329.
PTV/PTVBKF Injector: Microseal Valve Installation
a.
Place the Microseal valve in the injector head assembly.
b. Push down gently to seat the Microseal valve onto the rim of the septum holder.
c.
Screw the Microseal valve cap completely. Microseal valve overtightening and sealing
flange stressing is not mechanically allowed.
8. Close the module flap cover.
9. If present, move the autosampler towards the module to restore the original alignment.
10. Turn the carrier gas on.
11. Set the normal injector, detector, and GC working conditions.
Microseal Valve
Microseal Valve Cap
Injector Head Assembly