8
Installing Optional Kits
Performing the Dual FPD Detector Configuration
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
413
8. Remove the filter a support and the flange assembly from the cell of the secondary FPD
detector.
a.
Loosen the three smaller Allen screws that fix the flange to the detector body and
remove it with the spacer that should remain inserted in the flange See
.
Figure 425.
Filter Support and Flange Assembly Removal
b. Now you could access the flame-side heat shield, and the relevant graphite seal.
Loosen the Allen screws that fix the mirror plug and remove it. See
.
Figure 426.
Mirror Plug Removal
CAUTION
If the filter installed is equipped with the spacer (see the kits PN 19050785 and
PN 19050786), first remove the spacer, and then the filter.
Interferential Filter
Spacer
Fixing Screws