Installing Optional Kits
Installing the Merlin Microseal High Pressure Valve Kit
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
329
Figure 327.
SSL/SSLBKF and HeS-S/SL Injector: Microseal Valve Installation
a.
Place the Microseal valve in the septum holder.
b. Push down gently to seat the Microseal valve onto the rim of the septum holder.
c.
Screw the Microseal valve cap completely. Microseal valve overtightening and sealing
flange stressing is not mechanically allowed.
8. Close the module flap cover.
9. If present, move the autosampler towards the module to restore the original alignment.
10. Turn the carrier gas on.
11. Set the normal injector, detector, and GC working conditions.
To install the Merlin Microseal Valve kit on the PTV/PTVBKF injector
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
4. Put the autosampler away if present.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Septum Holder
Microseal Valve
Microseal Valve Cap