1
Installation
Installing the Autosampler
34
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Connecting the Autosampler
There are two possible cabling patterns for connecting the TRACE 1300/TRACE 1310 to a
single autosampler:
• The GC and a TriPlus, AI/AS 1310, or AI/AS 3000 II autosampler are controlled via a
single serial cable connected. In this case, a RS-232 connection is also needed between the
GC and the autosampler.
• In the Chromatography Data System (CDS)
Serial Port
configuration window, this
option is termed
Through GC
. Through GC is the default connection. See
Figure 43.
Through GC Connection
• The GC is connected to the Chromatography Data System (CDS) via LAN cable.
The autosampler is connected to the GC via RS-232 serial line and handshake cables.
See
Figure 44.
Separated Connections
Autosampler
Handshake Cable
Serial Cable
LAN Cable
Serial Cable
Handshake Cable