Performing Routine Maintenance
Shutting Down the TRACE 1300/TRACE 1310
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
71
Shutting Down the TRACE 1300/TRACE 1310
To shut down the GC
1. Cool down the GC.
2. If you do not plan to replace the column or perform maintenance on the GC, you do not
have to lower the injector temperature.
3. Turn off the carrier gas supply at the tank.
4. Push down the power switch (breaker), located at the back of the instrument, to the
position
O
(down). See
.
Figure 73.
GC Power Off
5. If external modules are present, unplug the power cable from the AC Input connector of
each module, and from the wall outlet.
6. If present, power-off the autosampler by switching off the main power switch, or by
unplugging the power cable from the AC input connector, and from the wall outlet.
7. Power-off all the remaining instruments.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Power Switch Off Position
AC Input Connector