4
Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
176
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Installing the FPD Detector
Before proceeding the installation of the FPD detector assembly, make sure that the FPD
control module is properly installed into the back detector housing.
To install the FPD detector assembly on the FPD detector base body
1. Place the jet into the detector base body housing and tighten it. Ensure the jet is perfectly
vertically aligned to avoid damage.
2. Place the FPD on the detector base body, ensuring that the aluminium ring has been
inserted in the correct position. Tighten the fixing nut by using the FPD fixing tool.
See
Paper towels
Forceps or tweezers
FPD Jet (if necessary)
Interferential filter for sulphur PN 28107000 or PN 19050785, (if necessary)
Interferential filter for phosphorous PN 28107100 or PN 19050785, (if necessary)
Interferential filter for tin PN 28107001 (optional), (if necessary)
FPD Ignition Glow-plug
T6 Torxhead key
T10 Torxhead screwdriver
5 mm wrench
5/16-in tube wrench
Note
For maintaining or replacing any other component not listed in this section, see
Chapter 7, “Detectors Advanced Maintenance.”
•
•
“Cleaning or Replacing the FPD Mirror Metal Plug”
•
“Cleaning or Replacing the FPD Filter-side Heat Shields”
•
“Cleaning or Replacing the FPD Flame-side Heat Shields”
•
“Replacing the FPD Photomultiplier Tube”
Materials needed to maintain a FPD detector