1
Installation
Installing the Injector and Detector Modules
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
5
a.
Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew the two captive fixing screws.
c.
Keeping the dummy module flap cover open, lift up the module from its seat in the
injector housing. Place the dummy module on a clean surface.
d. Remove the gas block plug from the gas connection by unscrewing its fixing screw
using a T20 Torxhead screwdriver.
Figure 2.
Injector Gas Block Plug
2. Plug the injector module into the main frame.
a.
Open the module flap cover.
b. Keeping the module flap cover open, place it in its seat. Be sure to insert the 25-pin
male connector on the bottom of the module into the 25-pin female connector on
the injector seat of the injector housing.
c.
Use a T20 Torxhead screwdriver to tighten the three captive fixing screws evenly and
carefully without overtightening.
Gas Connection Without Gas Block Plug
Gas Block Plug on Injector Seat
O-ring
WARNING
Make sure the o-ring is placed into its seat on the gas connection.
See
.
Do not install the module if the o-ring is missing
.
CAUTION
To maintain the correct alignment the screws must be tightened in turn, and
each screw must be tightened only a small amount before moving to the next screw.
Repeat until all are secure.