Detectors Advanced Maintenance
Measuring the FID Gas Flows
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
269
Figure 247.
FID Gases Measuring Tool Assembling
d. Place the flowmeter tubing over the male hose adapter of the measuring tool to begin
measuring flows. See
Figure 248.
Measuring Tool to Flowmeter Connection
12. Power on the GC.
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker) to the position I.
13. Measure the gas flow rate.
a.
Open the GC user interface.
b. Turn the
Hydrogen
flow On.
i.
Measure the gas flow and verify that the actual flow rate corresponds to the
setpoint value.