10
Adding Systems
Adding the PTV and PTVBKF Cryo System
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
517
c.
Insert the coolant tube into the insertion hole, and fix it using the screw previously
removed. See
Figure 560.
Installation of the Coolant Tubes for Carbon Dioxide and Liquid Nitrogen
d. Guide the coolant tube along the GC top cover up to reach the solenoid valve
assemble on the back of the GC. Bend the tube if necessary.
11. Connect the coolant tube to the solenoid valve assemble.
a.
Connect the coolant tube to the solenoid valve assembly using the proper nut and
ferrule. Use a 7/16-in. wrench for tightening the fittings. See
.
Coolant Tube for Carbon Dioxide
Coolant Tube for Liquid Nitrogen