8
Installing Optional Kits
Installing a FTIR Make-up Module
398
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
6. Remove the dummy module from the position where the FTIR module will be installed.
a.
Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew and remove the two captive fixing
screws.
c.
Keeping the dummy module flap cover open, lift up the module from its seat in the
detector housing. Place the dummy module on a clean surface.
d. Remove the gas block plug from the gas connections by unscrewing its fixing screw
using a T20 Torxhead screwdriver. See
.
Figure 406.
Detector Gas Block Plug
7. Open the front door of the GC.
8. Plug the FTIR Make-up module into the main frame.
a.
Open the module flap cover.
Gas Connections Without Gas Block Plug
Gas Block Plug on Detector Seat
O-rings
WARNING
Make sure all the four o-rings are placed into their seats on the gas
connection. See
.
Do not install the module if the o-rings are missing.