6
Injectors Advanced Maintenance
Cleaning the SSL Injector Body
236
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 220.
SSL Injector Body Cleaning
11. Clean the injector body.
a.
Put the injector body into an ultrasonic cleaner, filled with a methanol/acetone
mixture (1:1).
b. Sonicate the injector body for about half an hour.
c.
Using tweezers, remove the injector body from the bath, and dry it with an inert gas.
12. Reinstall the injector body.
a.
Reinstall and fix the injector body into its housing by screwing the two fixing screws.
b. Using tweezers, replace both the internal and external body head o-rings.
Note
Do not remove the carrier, split and purge lines o-rings.
SSL Injector Body Fixing Screws
SSL Injector Body
Carrier Line O-ring
Split Line O-ring
Purge Line O-ring
Body Head Internal O-ring
Body Head External O-ring