1
Installation
Column Installation Requirements
54
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 61.
SSL Injector Bottom Parts Removal
a.
Using the 1/2-in. wrench provided, unscrew the retaining nut with the base seal and
the washer from the bottom of the injector.
6. Install the adapter on the bottom the injector. See
.
Figure 62.
Adapter Installation on SSL Injector Bottom
a.
Fingertighten the retaining nut of the adapter until it start to grip the bottom of the
injector.
b. Use the 1/2-in. wrench to firmly tighten the retaining nut of the adapter with the
base seal and the washer from the bottom of the injector.
7. Remove the bottom parts of the FID, NPD, ECD, TCD, or FPD detector. See
Base Seal
Washer
Retaining Nut
SSL Injector Bottom
CAUTION
Make sure that the liner does not come from the bottom of the injector.
Save the bottom parts of the SSL injector in a safe place because will be reused when you
restore the original configuration.
Retaining Nut
Column Adapter
Nut
Silver Washer
Graphpacl SSL
Bottom
SSL Washer
Inlet