10
Adding Systems
Adding an Auxiliary Gas System
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
539
f.
Repeat
until all the components are connected to the auxiliary gas interface.
See
Figure 591.
Gas Tubes Connection to the Auxiliary Gas Interface - Internal View(
6. Connect the Supply Gas to the Auxiliary Gas Module.
a.
Connect the gas line to the corresponding inlet port of interest using the appropriate
nut and ferrules. Use a 7/16-in. wrench for tightening the fittings.
b. Repeat
until all the gas lines of interest are connected to the corresponding
inlet port on the auxiliary gas module. See
.
Figure 592.
Gas Line Connection to the Auxiliary Gas Interface
Note
Use the 1/8-in. Swagelok fittings provided on the gas inlet ports to connect the
gas lines.