Installing Optional Kits
Performing the Dual FPD Detector Configuration
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
417
13. Install the PT 100 simulator, provided with kit PN 19050783, on the detector base of the
front detector control module. See
.
Figure 434.
PT100 Simulator Installation
14. Install the Dual FPD on the detector base of the detector control module placed into the
back site of the detector housing. See
Figure 435.
Dual FPD Installation
a.
Place the detector on its base body, ensuring that the aluminium ring has been
inserted in the correct position, then tighten the fixing nut.
15. Connect the cables to the FPD detectors.
IMPORTANT
Save the jet and the aluminium ring in a safe place because will be reused
when you restore the original configuration of the FPD modules.
PT100 Simulator provided
with the kit PN 19050783