4
Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
189
7. Remove the covers of the FPD detector body.
a.
Using a 5/16-in tube wrench, remove the two screws on the top of the detector body,
and the front and back screws on the lower part of the detector body. See
Figure 176.
Detector Body Dismounting (1)
b. Remove the left and right covers of the FPD detector body and the insulating
material. The glow-plug is now accessible. See
.
Figure 177.
Detector Body Dismounting (2)
8. Replace the defective glow-plug. See
.
Note
Do not lose the aluminium ring inserted between the detector head and the
base body.
CAUTION
Removing the covers, pay attention to the heater element and the temperature
probe.
Top Screws
Right Cover
Front Screw
Back Screw
Glow-plug Cable
Terminal body
Right Cover
Left Cover
Insulating
Material
Insulating
Material