Detectors Advanced Maintenance
Cleaning or Replacing the FPD Flame-side Heat Shields
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
311
8. Remove the filter support.
a.
Loosen the three Allen screws that fix the filter support to the flange of the detector
body and remove it. See
Figure 302.
Filter Support Removal
9. Remove the filter-side heat shield.
a.
Loosen the three Allen screws that fix the flange to the detector body and remove it
with the spacer that should remain inserted in the flange. Now you could access the
flame-side heat shield, and the relevant graphite seal. See
Figure 303.
Filter-side Heat Shield Removal (1)
CAUTION
If the filter installed is equipped with the spacer (see the kits PN 19050785 and
PN 19050786), first remove the spacer, and then the filter.
Interferential Filter
Spacer
Flange
Filter Support