Injectors Advanced Maintenance
Cleaning the PTV Injector Head Assembly
252
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
18. Open the gas supplies.
19. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
20. Power on the GC.
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker) to the position I.
21. Set the normal oven, injectors, and detectors working conditions.