6
Injectors Advanced Maintenance
Cleaning the PTVBKF Injector Head Assembly
254
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
a.
Using the slotted stubby driver provided, unscrew and remove the liner cap.
b. Use tweezers to remove the liner with the liner seal (o-ring) from the injector.
Figure 233.
PTVBKF Top Parts and Liner Removal
10. Remove the fan assembly.
The fan is fixed on a support. Do not remove the fan from its support, but remove the
assembly by proceeding as follows:
a.
Using tweezers, and a T10 Torxhead screwdriver, unscrew and remove the left fixing
screw of the support, then unscrew and remove the right fixing screw. See
CAUTION
Be careful not to break the glass liner when removing it. Glass splinters
might fall into the lower part of the vaporization chamber. If the glass liner
breaks, follow the instructions for
“Replacing the PTVBKF Broken Liner”
Septum Cap
Septum
Liner Cap
Liner Seal (O-ring)
Liner
Injector Head Assembly