1
Installation
Connecting the Oven Cryogenic System
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
15
3. Connect the cryo valve to the
Aux Temperature/Cryo Module
.
a.
Connect the cryo solenoid valve to the 2-pin connector marked
Cryo Valves - Oven
using the cable provided. See
Figure 15.
Cryo Valves: Oven
b. Using the cable provided, connect the 15-pin female connector marked
GC Bus
on
the module to a
Bus
interface on the back of the GC.
c.
Plug the power cable to the AC Input connector on the front of the module and to
the wall outlet. The LED marked
On
lights up after the GC power on.
IMPORTANT
The Auxiliary Temperature/Cryo must be powered at the same line voltage
of the main GC system.
Note
For further details regarding the installation of the Aux Temperature/Cryo module,
“Adding an Aux Temperature/Cryo Module”
Oven Cryo Valve