GC Main Frame Advanced Maintenance
Removing/Replacing the GC Right Side Panel
200
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
4. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
5. Remove the AI/AS autosampler, if present.
6. Remove the panel.
a.
Open the front door of the GC. Using a T20 Torxhead screwdriver, unscrew the
right side panel screw from the interior front panel. See
.
Figure 186.
Right Panel Fixing Screw
b. Slide the panel towards the back of the instrument up to the stop.
c.
Remove the panel pulling it outwards being aware that the ground wire is attached to
the panel. See