3
Performing Injectors Routine Maintenance
Maintaining a Split/Splitless Injector with Backflush (SSLBKF)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
95
b. Using tweezers, remove the septum from the septum holder.
c.
Avoid touching the septum with your fingers. Insert a new septum into the septum
holder using tweezers.
d. Screw and tighten the septum cap to finger-tight.
7. Close the module flap cover.
8. If present, move the autosampler towards the module to restore the original alignment.
9. Turn the carrier gas on.
10. Set the normal injector, detector, and GC working conditions.
Cleaning or Replacing the SSLBKF Glass Liner
To clean or replace the glass liner
Figure 85.
SSLBKF Injector: Glass Liner Replacement
CAUTION
Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
Septum Cap
Ring Nut
Septum Holder/Liner Cap with Septum
Liner Seal (O-ring)
Liner