10
Adding Systems
Adding an Auxiliary Gas System
524
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
The module should be installed into a free external module housing provided on the back of
the GC. See
.
Figure 567.
External Modules Housing
Preliminary Operations
Before starting, the following preliminary operation must be carried out.
1. Put the GC in standby condition.
2. Cool the oven, injectors, and detectors to room temperature.
3. Close the gas supplies.
4. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector on the back of the GC, and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
External Modules Housing
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.