Installing Optional Kits
Installing the NoVent Microfluidics
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
379
11. Install the No-Vent module.
a. Open the module flap cover.
b. Keeping the module flap cover open, place the module in its seat. Be sure to insert the
25-pin male connector, on the bottom of the module, into the 25-pin female
connector on the detector seat of the detector housing.
c.
Use a T20 screwdriver to tighten the three captive fixing screws without
overtightening.
d. Close the module flap cover.
12. Power on the GC.
13. Switch the module on so that the LED light glows green and start the carrier gas flowing
to purge the line.
Installing the Mounting Bracket
Install the mounting bracket near the front of the GC oven on the right-hand side. This will
keep the mounting bracket out of the way of the column.
To install the mounting bracket
1. Remove the nut at the top end of the mounting bracket. This nut secures the two sections
of the mounting bracket. See
.
Figure 375.
Mounting Bracket Parts
2. Push the top of the mounting bracket down.
3. Line up the bottom of the mounting bracket into two holes on the bottom side of the GC
oven.
Note
Tighten the center screw first and then secure the side screws.
Note
This module requires a constant carrier gas pressure of 60 psig (410 kPa).
Note
If you already have a mounting bracket installed in your GC with either the SSL
backflush or PTV backflush system, skip this step.
Top
Bottom
Nut
Holder