8
Installing Optional Kits
Installing the NoVent Microfluidics
394
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 401.
Column Setup Options
5. Click
Apply
.
6. Click
Close
.
Using the Module
The GC does not recognize the presence of the NoVent microfluidics, so it is invisible to the
system. The module receives only the voltages for supplying the solenoid valve and the
On/Off light from the GC.
The normal condition of the module is
Off
. If the switch is in the
Off
position,
the Status
LED is
off
. See
.
Figure 402.
Module in Off Condition
Before replacing the column or performing maintenance on an injector module, flip the
switch to the
On
position. Ensure the mass spectrometer is in standby mode and the filament
is off. The status LED lights up as solid green indicating that the solenoid valve is activated.
The carrier gas flows into the mass spectrometer to avoid breaking vacuum. See
Post-column Settings
Status LED Off
Switch on Off Position