TM 11-6625-3145-14
Operating Instructions-318/338 Service
RMT Test. This test is available only with the 318S1/338S1. To begin this test press B while in the Diagnostics menu.
This test is for the SIO used by the remote mode. Refer to the Diagnostics menu or to the trouble-shooting tree in
for details. Figure 3-23 illustrates the display for the remote SIO test.
operation: Your equipment setup should match that in Figure 3-22 before you begin this test. To begin the test, press
START. The LOOP test and DISP ON/OFF functions are available here. The LOOP test has four features:
OFF, I/O, ERROR, and TEST. These can be selected by pressing SELECT, even while the test is in
progress. DISP can also be selected during testing. If DISP is OFF, no results will appear on the screen
other than results of the I/O LOOP test. To exit this test, wait until the test is complete, or press STOP, and
then press X. If errors are detected and DISP is ON, some of the following error codes will appear on the
screen.
Error codes : 96, 97, 98, 99, 9A, 9B, 9C, 9D, 9E, 9F, A0, A1, A2, A3, A4, A5, A6, A7, A8, A9, AA, AB, AC, AD, AE, AF, BO,
B1, B2 (Refer to
NOTE
This test may take up to 40 seconds to complete.
Figure 3-22. Equipment setup for remote test (RMT).
Figure 3-23. Display for remote tests (RMT).
3-16
Summary of Contents for 318
Page 119: ...318 VERIFICATION AND ADJUSTMENT PROCEDURES ...
Page 182: ...338 VERIFICATION AND ADJUSTMENT PROCEDURES ...
Page 253: ...318 ___________________ TROUBLESHOOTING TREES ...
Page 344: ...338 TROUBLESHOOTING TREES ...
Page 517: ...TM 11 6625 3145 14 318 338 4434 923 318 Block Diagram ...
Page 518: ...TM 11 6625 3145 14 318 338 4434 924 338 Block Diagram ...
Page 519: ...TM 11 6625 3145 14 318 338 4434 925 318 Acquisition Module Wiring Diagram ...
Page 520: ...TM 11 6625 3145 14 318 338 4434 926 318 338 Mainframe Wiring Diagram ...
Page 521: ...TM 11 6625 3145 14 318 338 4434 926 338 Acquisition Module Wiring Diagram ...
Page 522: ...TM 11 6625 3145 14 318 338 4434 928 Figure 9 1 318 A01 Input A Board Component Locations ...
Page 526: ...TM 11 6625 3145 14 ...
Page 528: ...TM 11 6625 3145 14 Figure 9 3 318 338 A03 ACQ Control Board Component Locations ...
Page 532: ...TM 11 6625 3145 14 ...
Page 536: ...TM 11 6625 3145 14 ...
Page 538: ...TM 11 6625 3145 14 ...
Page 539: ...TM 11 6625 3145 14 ...
Page 540: ...TM 11 6625 3145 14 ...
Page 541: ...TM 11 6625 3145 14 ...
Page 542: ...TM 11 6625 3145 14 Figure 9 8 318 338 A10 CRT Board Component Locations ...
Page 544: ...TM 11 6625 3145 14 Figure 9 9 318 338 A11 Inverter Board component Locations ...
Page 546: ...TM 11 6625 3145 14 Figure 9 10 318 338 A12 Regulator Board Component Locations ...
Page 551: ...TM 11 6625 3145 14 Figure 9 12 338 A01 Input A Board Component Locations ...
Page 553: ...TM 11 6625 3145 14 318 338 SERVICE ...
Page 554: ......
Page 555: ...PIN 058584 ...