![Tektronix 318 Technical Manual Download Page 129](http://html1.mh-extra.com/html/tektronix/318/318_technical-manual_1077905129.webp)
TM 11-6625-3145-14
Verification and Adjustment Procedures-318/338 Service
Threshold Level V2
a.
Set Threshold LEVEL V2 to +0.3 V.
b. Set both Pod A and Pod B equal to V2.
c. Press the START key. The trigger position is displayed on the screen will be <T=7W>, <T=127W>, or <T=247W>
depending on the trigger position setting in the Trigger menu.
d. Check that all data acquired are 0’s.
e. Set Threshold LEVEL V2 to -0.3 V.
f. Press the START key. The Trigger position displayed on the screen will be <T=7W>, <T=127W>, or <T-247W>
depending on the trigger position setting in the Trigger menu.
g. Check that all data acquired are l’s.
Threshold Level V3
a. Set Threshold LEVEL V1 to +2.0V and V2 to -1.5 V.
b. Check that the Threshold LEVEL V3 is -+0.25 V.
c. Set both Pod A and Pod B equal to V3.
d. Press the START key. The trigger position displayed on the screen will be <T=7W>, <T=127W>, or <T-247W>
depending on the trigger position setting in the Trigger menu.
e. Check that all data acquired are 0’s.
f. Set Threshold LEVEL V1 to -2.0 V and V2 to + 1.5 V.
g. Check that Threshold LEVEL V3 is -0.25 V.
h. Press the START key. The trigger position displayed on the screen will be <T=7W>, <T= 127W>, or <T=247W>
depending on the trigger position setting in the Trigger menu.
i. Check that all data acquired are l’s.
5-10
Summary of Contents for 318
Page 119: ...318 VERIFICATION AND ADJUSTMENT PROCEDURES ...
Page 182: ...338 VERIFICATION AND ADJUSTMENT PROCEDURES ...
Page 253: ...318 ___________________ TROUBLESHOOTING TREES ...
Page 344: ...338 TROUBLESHOOTING TREES ...
Page 517: ...TM 11 6625 3145 14 318 338 4434 923 318 Block Diagram ...
Page 518: ...TM 11 6625 3145 14 318 338 4434 924 338 Block Diagram ...
Page 519: ...TM 11 6625 3145 14 318 338 4434 925 318 Acquisition Module Wiring Diagram ...
Page 520: ...TM 11 6625 3145 14 318 338 4434 926 318 338 Mainframe Wiring Diagram ...
Page 521: ...TM 11 6625 3145 14 318 338 4434 926 338 Acquisition Module Wiring Diagram ...
Page 522: ...TM 11 6625 3145 14 318 338 4434 928 Figure 9 1 318 A01 Input A Board Component Locations ...
Page 526: ...TM 11 6625 3145 14 ...
Page 528: ...TM 11 6625 3145 14 Figure 9 3 318 338 A03 ACQ Control Board Component Locations ...
Page 532: ...TM 11 6625 3145 14 ...
Page 536: ...TM 11 6625 3145 14 ...
Page 538: ...TM 11 6625 3145 14 ...
Page 539: ...TM 11 6625 3145 14 ...
Page 540: ...TM 11 6625 3145 14 ...
Page 541: ...TM 11 6625 3145 14 ...
Page 542: ...TM 11 6625 3145 14 Figure 9 8 318 338 A10 CRT Board Component Locations ...
Page 544: ...TM 11 6625 3145 14 Figure 9 9 318 338 A11 Inverter Board component Locations ...
Page 546: ...TM 11 6625 3145 14 Figure 9 10 318 338 A12 Regulator Board Component Locations ...
Page 551: ...TM 11 6625 3145 14 Figure 9 12 338 A01 Input A Board Component Locations ...
Page 553: ...TM 11 6625 3145 14 318 338 SERVICE ...
Page 554: ......
Page 555: ...PIN 058584 ...