TM 11-6625-3145-14
Verification and Adjustment Procedures-318/338 Service
8.
Set serial data generator SETUP as follows;
CODE=ASCII, BAUD=2400, DUPLEX=FULL, SETUP=ASYNC, BITS/CHAR=8, PARITY= NONE, STOPBITS = 1,
TIMING= NORMAL..
9.
Load the following program into the serial data generator:
STEP
MNEMONIC
CODE
1
SEND
#1
2
JMP
#1
10. Load the following hexadecimal data in the serial data generator buffer # 1: E0, E0, E2, E4, E6, E8
11. Press the START key on the serial data generator.
12. Select the DATA DISPLAY STATE TABLE and press the START key.
13. When the 318 displays data on the crt, press the serial data generator STOP key.
14. Check that the 318 display matches the WORD SEQUENCE column on Table 5-12.
15. Set the 318 baud rate to 19.2K (DATA screen).
16. Set the serial data generator baud rate to 19.2K.
17. Set the 318 and the serial data generator controls as indicated in step 1 of Table 5-12 and repeat preceding steps of
Test 6 (11) through (14).
18. Finish steps 2 through 7 Table 5-12 as set up in step (17) of Test 6.
TABLE 5-12.
318 SERIAL STATE ANALYZER TEST STEPS
Serial Data
Step
318
Generator
Word Sequence
1
BITS/WORD = 8
BITS/CHAR = 8
E0, E0, E2, E4, E6. E8
2
BITS/WORD = 7
BITS/CHAR = 7
60, 60, 62, 64, 66, 68
3
BITS/WORD = 6
BITS/CHAR = 6
20, 20, 22, 24, 26, 28
4
BITS/WORD = 5
BITS/CHAR = 5
00, 00, 02, 04, 06, 08
5
BAUD RATE = EXT16
BITS/WORD = 6
BITS/CHAR = 6
20, 20, 22, 24, 26, 28
6
COMM. MODE = SYNC
BITS/WORD = 8
SYNC WORD 1 = 16
SYNC WORD 2 = 16
SETUP = SYNC
BITS/CHAR = 8
CLOCK = DERIVED
(INSERT DATA 16, 16 TO
TOP OF BUFFER #1)
E0, E0, E2, E4, E6, E8,
FF, FF, FF
7
BITS/WORD = 5
SYNC WORD 1 = AD
SYNC WORD 2 = 05
HUNT WORD = 08
BITS/CHAR = 5
00, 02,04,06, 08
5-62
Summary of Contents for 318
Page 119: ...318 VERIFICATION AND ADJUSTMENT PROCEDURES ...
Page 182: ...338 VERIFICATION AND ADJUSTMENT PROCEDURES ...
Page 253: ...318 ___________________ TROUBLESHOOTING TREES ...
Page 344: ...338 TROUBLESHOOTING TREES ...
Page 517: ...TM 11 6625 3145 14 318 338 4434 923 318 Block Diagram ...
Page 518: ...TM 11 6625 3145 14 318 338 4434 924 338 Block Diagram ...
Page 519: ...TM 11 6625 3145 14 318 338 4434 925 318 Acquisition Module Wiring Diagram ...
Page 520: ...TM 11 6625 3145 14 318 338 4434 926 318 338 Mainframe Wiring Diagram ...
Page 521: ...TM 11 6625 3145 14 318 338 4434 926 338 Acquisition Module Wiring Diagram ...
Page 522: ...TM 11 6625 3145 14 318 338 4434 928 Figure 9 1 318 A01 Input A Board Component Locations ...
Page 526: ...TM 11 6625 3145 14 ...
Page 528: ...TM 11 6625 3145 14 Figure 9 3 318 338 A03 ACQ Control Board Component Locations ...
Page 532: ...TM 11 6625 3145 14 ...
Page 536: ...TM 11 6625 3145 14 ...
Page 538: ...TM 11 6625 3145 14 ...
Page 539: ...TM 11 6625 3145 14 ...
Page 540: ...TM 11 6625 3145 14 ...
Page 541: ...TM 11 6625 3145 14 ...
Page 542: ...TM 11 6625 3145 14 Figure 9 8 318 338 A10 CRT Board Component Locations ...
Page 544: ...TM 11 6625 3145 14 Figure 9 9 318 338 A11 Inverter Board component Locations ...
Page 546: ...TM 11 6625 3145 14 Figure 9 10 318 338 A12 Regulator Board Component Locations ...
Page 551: ...TM 11 6625 3145 14 Figure 9 12 338 A01 Input A Board Component Locations ...
Page 553: ...TM 11 6625 3145 14 318 338 SERVICE ...
Page 554: ......
Page 555: ...PIN 058584 ...