S5105F Microwave Analyzer Test Results (Continued 2)
No.
Item
Unit
Standard requirement
Results
12
Residual response
(Preamplifier
On)
dBm
10MHz - 40GHz
≤-70
13
Overall absolute
amplitude accuracy
(attenuation 10dB,
input -10dBm)
dB
500MHz:
±2.00
2.5GHz:
±2.00
4.5GHz:
±2.00
6.5GHz:
±2.00
8.5GHz:
±2.00
10.5GHz:
±2.00
12.5GHz:
±2.00
14.5GHz:
±2.00
16.5GHz:
±2.00
18.5GHz:
±2.30
20.5GHz:
±2.30
22.5GHz:
±2.30
24.5GHz:
±2.30
27.0GHz:
±2.70
29.0GHz:
±2.70
31.0GHz:
±2.70
33.0GHz:
±2.70
35.0GHz:
±2.70
37.0GHz:
±2.70
39.0GHz:
±2.70
14
Maximum safe input
level
/
+27dBm (continuous wave, input attenuator is
automatically coupled)
15
VBW
/
Bandwidth range: 1Hz - 5MHz (1-3 sequence)
Vector network analyzer performance specification
16
Frequency range
and accuracy
/
50MHz - 40GHz:
1GHz:
1GHz±1000Hz
17
Power scope
/
Large, small and manual setting
18
Power step
dB
1
19
Port output power
dBm
-11 - +1
20
Power level
accuracy
dB
100MHz
±2.5
1GHz
±2.5
3GHz
±2.5
5GHz
±2.5
7GHz
±2.5
9GHz
±2.5
11GHz
±2.5
13GHz
±2.5
15GHz
±2.5
17GHz
±2.5