134
c) Press [Trace] → [Trace Math>] → [Save] to display only the trace corresponding to the saved data;
d) Press [Trace] → [Math >] → [Data & Mem] to simultaneously display the trace corresponding to the
measurement data and the stored data;
e) Press [Trace] → [Trace Math>] → [Data-Mem] to display the trace corresponding to the
measurement data minus the saved data;
f)
Press [Trace] → [Math >] → [Data/Mem] to display the corresponding trace that the measurement
data is divided by stored data;
g) Press [Trace] → [Math >] → [Data + Mem] to display the corresponding trace that measurement
data plus the stored data.
2
Limit line measurement and editing
a) Press [Trace] → [Limit >] to enter the limit function menu.
b) Press [Limit >] → [Limit Off On] to open the limit line display function. At this time, the limit line
composed of the frequency points just edited will be displayed on the screen, and the upper limit line
is marked by the green line;
c) Press [Limit>] → [Alarm Off/On] to enter the On status of the menu. If the limit line test fails, the
microwave analyzer sends out the “tick & tick” sound for alarming;
d) Press [Limit>] → [Clear] to clear the edited limit line displayed currently;
e) Press [Limit>] → [Edit>] → [Sel Point 1] to switch the limit point of current limit;
f)
Press [Limit>] → [Edit>] → [Pos] to set current limit point frequency, enter number via the figure key,
and select the frequency unit for completion;
g) Press [Limit >] → [Edit >] → [Value] to type the amplitude value of the frequency point with numeric
keys and select the amplitude unit to complete the input;
h) Press [Limit >] → [Edit >] → [Add] to add the limit point on the current limit line;
i)
Press [Limit>] → [Edit>] → [Delete] to delete the limit point displayed on current limit;
j)
Press [Limit>]→[Edit>] → [Clear] to clear all limit points saved
on current limit;
k) Press [Limit>] → [Mem] to type the name with the numeric keypad, and save the previously edited
limit line into the system. When you make the same measurement next time, manually select it by
pressing [Limit] → [Recall >] with the touch screen. Then, the recall of limit line is completed by
pressing [OK], which saves the user from repeating the editing work;
l)
Press [Limit>] → [Recall>], manually select the limit line file with the touch screen and press [OK]
to complete the recall. If the number of limit line storage files is large, press [Limit>] → [Recall>] →
[Page UP], [Page UP] and find the limit line file to be recalled by page, and press [Limit>] → [Recall>]
→ [Delete], [Del All] to delete the single and all currently selected and stored limit line files.
Section 7 Marker Search and Settings
This Section describes the peak search and the related use method of the marker in the antenna and
cable test mode of S5105D/E/F microwave analyzer. There is no marker operation in power
monitoring and signal generation mode.
1
Marker functions
The S5105D/E/F microwave analyzer provides up to 8 independent markers for each trace in the antenna
and cable test mode for reading out measurements. Each marker can be set to the normal mode or
△
mode with the marker off and all off function. Follow the steps below to use the marker:
a) Press
【
Mkr
】
→ [Select Mkr [1] ] → [Mkr 1], [Mkr 2], [Mkr 3], [Mkr 4], [Mkr 5], [Mkr 6] or [More Mkr]
→ [Mkr 7], [Mkr 8], select the currently displayed marker;
b) At this time, the marker is in the normal mode, which can be moved via the rotary knob, step key or
figure key;