Tel: 886.909 602 109 Email: [email protected]
187
S5105F Microwave Analyzer Test Results (Continued 1)
No.
Item
Unit
Standard requirement
Results
7
RBW Accuracy
/
5MHz
±15.0%
3MHz
±10.0%
1MHz
±10.0%
300kHz
±10.0%
100kHz
±10.0%
30kHz
±10.0%
10kHz
±10.0%
3kHz
±10.0%
8
Sideband noise (carrier
frequency 1GHz)
dBc/Hz
-30kHz
≤-102.0
+30kHz
≤-102.0
+100kHz
≤-99.0
-100kHz
≤-99.0
+1MHz
≤-110.0
-1MHz
≤-110.0
9
Explicit average noise
level
dBm
Preamplifier
On
10MHz - 4.5GHz
≤-151.0
4.5GHz - 7GHz
≤-147.0
7GHz - 13GHz
≤-145.0
13GHz - 18GHz
≤-140.0
18GHz - 26.5GHz
≤-138.0
26.5GHz - 40GHz
≤-135.0
dBm
Preamplifier
Off
10MHz - 4.5GHz
≤-135.0
4.5GHz - 7GHz
≤-131.0
7GHz - 13GHz
≤-127.0
13GHz - 18GHz
≤-120.0
18GHz - 26.5GHz
≤-116.0
26.5GHz - 40GHz
≤-113.0
10
Second harmonic
distortion
dBc
0.9GHz
≤-60.0
6GHz
≤-60.0
11
Mirror,
multiple and
out-of-band
responses
2GHz
dBc
Image frequency 267.5MHz
≤-70.0
Image frequency 9,267.5MHz
≤-70.0
5GHz
dBc
Image frequency 6,732.5MHz
≤-80.0
Image frequency 24,732.5MHz
≤-80.0
7.6GHz
dBc
Image frequency 5,867.5MHz
≤-65.0
Image frequency 14,867.5MHz
≤-65.0
12.4GHz
dBc
Image frequency 10,667.5MHz
≤-55.0
Image frequency 19,667.5MHz
≤-55.0
19.4GHz
dBc
Image frequency 17,667.5MHz
≤-65.0
Image frequency 12,132.5MHz
≤-65.0
28.5GHz
dBc
Image frequency 26,767.5MHz
≤-65.0
Image frequency 21,232.5MHz
≤-65.0