b) Press [GPS Off/On] to switch the GPS status, and enable the GPS function.
c) Press [GPS Info] to view detailed GPS information.
7
System language selection
a) Press [System]→[Language] to enter the language selection interface of the system.
b) Press [Simplified Chinese] or [English] to set current interface operation language.
8
Set date and time
a) Press [System]→[Set]→[Date&Time] to pop up the date setting dialog box.
b) Click corresponding zone in the calendar area to set the date.
c) Click corresponding zone in the time area to enter numerical value for time setting.
d) Press
【
OK
】
to complete the setting and close the dialog box.
9
Network configuration
It is used to set the microwave analyzer network, with specific steps shown below:
a) Press [System]→[Set]→[LAN] to pop up the network configuration dialog box.
b) Click corresponding zone to enter numerical value for setting.
c) Press
【
OK
】
to complete the setting and close the dialog box.
10 Software update
The system software of the S5105D/E/F microwave analyzer can be updated via the U disk, with
steps shown below:
a) Copy files necessary for software update via the computer to the root directory of the U disk, and
insert the U disk to the A-type USB port of the microwave analyzer.
b) Press [System]→[Set]→[Update] to pop up the software update interface.
c) Press
【
OK
】
for software update. After that, a dialog box will pop up for returning to the original test
procedure.
d) Press
【
Reset
】
to restart the microwave analyzer and complete software update.
NOTE:
This function can be used only by the factory debugging personnel or
technical supporting personnel rather than users. Otherwise, the
microwave analyzer can be damaged.
11 Factory debugging
The S5105D/E/F microwave analyzer is provided with a program entry for factory debugging.
a) Press [System]→[Configuration]→[Manage] to pop up the “Enter Password” dialog box, and enter
the factory debugging menu after that.
NOTE:
This function can be used only for factory debugging, and users cannot
try to crack the password. Otherwise, the microwave analyzer can be
damaged.
12 Set frequency reference
The S5105D/E/F microwave analyzer is provided with both the internal frequency reference and
external one. The external reference frequency must be 10MHz±100Hz and amplitude of 0dBm
(limit range:
-2dBm - +10dBm). The external reference frequency must be entered from the “10MHz input/output” port
on the cover plate.
a) Press [System]→[Page Down] to set the frequency reference, which is the internal frequency
reference by default.
!
!