Tel: 886.909 602 109 Email: [email protected]
180
S5105D Microwave Analyzer Test Results (Continued 3)
No.
Item
Unit
Standard requirement
Results
22
Corrected source
matching
dB
≥37 2MHz -
500MHz
1 port (S11)
2 port (S22)
≥30
500MHz - 9GHz
1 port (S11)
2 port (S22)
≥28
9GHz - 18GHz
1 port (S11)
2 port (S22)
23
Reflected tracking
dB
±0.10
2MHz - 500MHz
1 port (S11)
2 port (S22)
±0.13
500MHz - 9GHz
1 port (S11)
2 port (S22)
±0.14
9GHz - 18GHz
1 port (S11)
2 port (S22)
24
Transmission tracking
dB
±0.25
2MHz - 500MHz
1 port (S12)
2 port (S21)
±0.29
500MHz - 9GHz
1 port (S12)
2 port (S21)
±0.33
9GHz - 18GHz
1 port (S12)
2 port (S21)
25
Dynamic Range
dB
≥85.0 2MHz -
18GHz
1 port (S12)
2 port (S21)
26
interface
RF interface
/
RF input interface (corresponding to the interface
form required by the product model)
Communication
Interface
/
USB interface: type A, two
/
USB interface: mini type, one
/
LAN interface: RJ45 type
Frequency
reference
/
Reference input/output, BNC female type
Other interface
/
GPS antenna interface: BNC female type
/
External trigger input interface: BNC female type
27
Safety
/
Dielectric Strength requires AC 1,500V, 10mA/min;
no breakdown, no arcing.
/
Voltage 242 V, leakage current ≤3.5 mA, 1min.
/
The insulation resistance between the power input
end and the casing should not be less than 100 MΩ
under standard atmospheric conditions for testing.
/
The insulation resistance between the power input
end and the casing should not be less than 2 MΩ in
humid environment.
Description
1. “√”indicates normal function or meet the requirements; “X” indicates that the function is abnormal or does
not meet the requirements;
2. "/" indicates that there is no such test item on this device.
Comprehensive judgment: Pass □ Fail □