148
Show a menu displaying the calibration function.
[Cal Off On]
Turn the error correction on or off, and the calibration can only be turned on when the corresponding
calibration is completed.
[Cal Kit]
Select the type of calibration kit for mechanical calibration.
[31123]
Select a 2.4mm calibration kit.
[31121]
Select a 3.5mm calibration kit.
[31101]
Select the 50Ω coaxial calibration kit.
[M Cal]
Select calibration method
[OPEN]
Select the calibration standard of OPEN.
[SHORT]
Select the calibration standard of SHORT.
[LOAD]
Select the calibration standard of LOAD.
[Done]
After completing the measurement of the calibration standards for the open device, short device and
load, follow this menu item to complete the calibration of the calibration standard and the correction
factor, and automatically turn on the calibration switch.
[Cal Cont>]
Complete the remaining calibration steps.
【
Measure
】
[Return Loss]
Set the measurement format to return loss.
[Cable Loss]
Set the cable loss for DTF measurement, and this parameter is valid when the correction mode is
manual. After the setting tab is popped up, enter the value to be set with numeric keys, press the [OK]
key to complete the parameter setting, or adjust it by the step key or rotary knob.
[VSWR]
Set the measurement format to VSWR.
[DTF Return Loss]
Set the measurement format to
DTF
return loss.
[DTF VSWR]
Set the measurement format to
DTF
VSWR.
[Phase]
Set the measurement format to phase.
[Smith>]