Chapter VII Power Monitoring Mode (Option)
The power monitoring mode (hereinafter referred to as "PM") is an optional test function option for the
S5105D/E/F microwave analyzer, and the measured frequency range covers 100kHz - 18GHz
(S5105D), 100kHz - 26.5GHz ( S5105E), 100kHz - 40GHz (S5105F) . This mode eliminates the
necessity of connecting an external power probe, and can easily test the power of a continuous wave
signal under a certain bandwidth, and the channel bandwidth can be set to a larger width to simulate
the average power meter measurement.
Section 1 Basic Operations
This Section will briefly introduce the power monitoring mode measurement function and measurement
method of the S5105D/E/F microwave analyzer, so that the first-time users can have a general
understanding of the test operation process and perform basic measurement after reading this section.
1
Select modeSelect mode
Connect the signal to be tested to the spectrum input port of the
【
Mode
】
Analyzer through the cable.
After preheating, press→ [Power Monitor], select the power monitoring mode, and press [OK] to
complete the mode setting.
2
Set the reference level
Press the
【
Amplitude
】
and the setting options related to the reference are displayed on the right side
of the screen, press [Ref Level 0.00dBm] → [0] → [dBm] , set the reference level to 0dBm.
3
Set Frequency Information
Press
【
Frep
】
→ [Center Freq] → [1] → [GHz], set the center frequency to 1GHz, and press
【
Frep
】
→ [Sweep >] → [1] [0] [MHz] to set the span to 10MHz.
Section 2 Frequency Setting
This Section describes the detailed setting method of frequency information in the power monitoring
mode of the S5105D/E/F microwave analyzer, and how to open the frequency setting soft menu
through the
【
Frep
】
button on the front panel.
1
Set Sweep Frequency
There are two ways to set the frequency in the power monitoring mode of the S5105D/E/F microwave
analyzer:
Set start and stop frequencies
a)
Press
【
Frep
】
→ [Start Freq];
b)
Enter the starting frequency value with the numeric keys and stop the input by pressing
corresponding frequency unit on the touch screen.
c)
Press [Stop Freq];
d)
Enter the stop frequency value with the figure keys and stop the input by pressing corresponding
frequency unit on the touch screen.
Set center frequency and span
a)
Press
【
Frep
】
→ [Center Freq];
b)
Enter the center frequency value with the figure keys and stop the input by pressing corresponding
frequency unit on the touch screen.
c)
Press [Span];
d)
Enter the Span value with numeric keys and stop the input by pressing corresponding frequency
unit on the touch screen.