Tel: 886.909 602 109 Email: [email protected]
190
S5105F Microwave Analyzer Test Results (Continued 4)
No.
Item
Unit
Standard requirement
Results
24
Transmission tracking
dB
±0.25
50MHz - 500MHz
1 port
(S12)
2 port
(S21)
±0.29
500MHz - 9GHz
1 port
(S12)
2 port
(S21)
±0.33
9GHz - 18GHz
1 port
(S12)
2 port
(S21)
±0.35
18GHz - 26.5GHz
1 port
(S12)
2 port
(S21)
±0.40
26.5GHz - 33GHz
1 port
(S12)
2 port
(S21)
±0.50
33GHz - 40GHz
1 port
(S12)
2 port
(S21)
25
Dynamic Range
dB
≥85.0
50MHz - 18GHz
1 port
(S12)
2 port
(S21)
≥80.0
18GHz - 26.5GHz
1 port
(S12)
2 port
(S21)
≥75.0
26.5GHz - 33GHz
1 port
(S12)
2 port
(S21)
≥65.0
33GHz - 40GHz
1 port
(S12)
2 port
(S21)
26
interface
RF interface
/
2.4mm positive
Communication
Interface
/
USB interface: type A, two
/
USB interface: mini type, one
/
LAN interface: RJ45 type
Frequency
reference
/
Reference input/output, BNC female type
Other interface
/
GPS antenna interface: BNC female type
/
External trigger input interface: BNC female type
27
Safety
/
Dielectric Strength requires AC 1500V, 10mA/min;
no breakdown, no arcing.
/
Voltage 242 V, leakage current ≤3.5 mA, 1min.
/
The insulation resistance between the power input
end and the casing should not be less than 100 MΩ
under standard atmospheric conditions for testing.
/
The insulation resistance between the power input
end and the casing should not be less than 2 MΩ in
humid environment.
Description
1. “√”indicates normal function or meet the requirements; “X” indicates that the function is abnormal or does
not meet the requirements;
2. A “/” indicates that the unit does not have this test item.
Comprehensive judgment: Pass
□
Fail
□