20
on top of the grid, and the start/stop frequencies displayed at the bottom. The trace is displayed in
coordinate grid in the spectrum analyzer mode. Scale values and mark information are displayed in
the left and upper right of the grid respectively. Such information as the resolution bandwidth, video
bandwidth and sweep time are displayed on top of the grid, and the center frequency and span
displayed at the bottom. In the power monitoring mode, the power readings measured are in two
reading modes, namely, the dials and numbers. The center frequency is displayed at the bottom of
the grid. In the vector voltage measurement mode, it displays such information as frequency of
current frequency point, calibration on/off state, measuring amplitude and phase. In the USB Power
Meter mode, the power readings measured are in two reading modes, namely, the dials and
numbers. The center frequency is displayed at the bottom of the grid. The contents displayed vary
with the measurement mode.
Submenu zone
It is used to display various-level submenus. Function hardkeys on the front panel and main menu
keys at the bottom of the display zone are all corresponding to the first-level submenus, with some
of the submenu keys corresponding to the second-level submenu. Conduct relevant instrument
settings via the submenu.
System status bar
It locates on top of the screen to display such information as the system date, time, current
measurement mode, current power supply type and battery capacity from left to right in sequence.
The signs of the power supply type and battery status indicator showed vary with the situation that
whether there is external power supply or the battery capacity. Signs corresponding to various
power supply statuses are shown below:
a) If the microwave analyzer is not provided with battery and adopts external power supply, it shows
.
b) If the microwave analyzer is provided with battery of full capacity (100%) and without or with
external power supply adapter connection, it shows
or
respectively. That is,
there are 1-5 green blockages in the white box, with each blockage accounting for 20% of the
capacity. If the microwave analyzer is not connected with an external adapter with battery capacity
of less than 100%, it shows a sign of less than 5 blockages, for example,
. The blockages
will decrease gradually with the battery capacity decrease during operation.
c) If the microwave analyzer is provided with battery and connected to external power supply adapter,
with battery capacity of less than 100% and battery being charged, it shows
, that is, the
number of the green blockages is changed from 1 to 5 in cycle, and kept at 5 after full charging.
d) If the microwave analyzer is not connected with external power supply with capacity of less than
10% and battery in the undervoltage status, it shows
, that is, there is a red blockage in
the white box. In this case, charge the battery in time. When the capacity is less than 10%, the
microwave analyzer can run for another 10 min. Please save the measurement results in time
because it will be shut down automatically after the power is used up.
Main menu zone
The function hardkeys on the main menu zone and the front panel are at the same level. For specific
measurement mode, the main menu generally consists of a set of unique menu keys for
supplementing the function hardkeys.
Section 3 System setting
The section introduces system-related settings in various modes of the S5105D/E/F microwave analyzer.
1
View version serial number
Each S5105D/E/F microwave analyzer is provided with a unique and fixed serial number and
internal software version information, which is set by the manufacturer upon delivery, cannot be
changed the user, and can be viewed by the following steps: