112
[Trace ]
A soft menu related to the trace pops up, including [Trace 1 2 3], [Clear Write], [Max Hold], [Min Hold],
[View], [Blank], [Limit].
[Trace 1 2 3]
Switch the current trace, the comprehensive analyzer spectrum analyzer mode provides 1, 2, 3 traces,
the selected trace number is underlined, and the status menu item where the trace is located is also
switched to the current trace state.
[Clear Write]
All data of the previously displayed trace is refreshed and the signal received in the sweep state is
continuously displayed.
[Max Hold]
The point on the selected trace is kept at its maximum value and updated with the new maximum
detected in each sweep. If the current detection mode is automatic, the detection mode will switch to
positive peak detection.
[Min Hold]
The point on the selected trace is kept at its minimum value and updated with the new minimum detected
in each sweep. If the current detection mode is automatic, the detection mode switches to negative peak
detection.
[View]
Keep and display the amplitude data of the selected trace, but it is not updated when sweeping.
[Blank]
The signal is only performed on the background and not displayed on the screen.
[Limit]
The limit line may be edited or displayed and tested to see if it is passed.
[Cur Limit Up Low]
Switch the currently displayed limit to the upper or lower limit.
[Alarm Off On]
Warning sounds when the limit line fails.
[Limit Off On]
Test switch for limit line.
[Sel Point]
Select the limit point of the current edit.
[Edit]
Enter the edit menu of the upper limit line.
[Pos]
Set the horizontal axis for certain limit point of the limit line.
[Value]
Set the amplitude of a certain frequency point of the limit line.
[Add]
Add a frequency point.
[Delete]
Delete a frequency point.
[Clear]