Operation Manual
SECTION 6
Communication with External Devices
ZERO
Function 1......Executes Force-zero.
ZERO 0 <Output ><Offset value>
•
Specify an output between 0 and 8.
•
Set the offset value used for force-zero execution.Set the value to be added to 0 in
the range from -999.999999 to 999.999999.
If this setting is omitted, the offset value that was set previously will be used.
Command executed correctly
: OK
Command not executed correctly
: ER
Example:
When the force-zero function is executed, the zero value is determined using the latest
measurement value even if the hold output or logging output is specified.
Output #
Meaning
0
OUT0
1
OUT1
2
OUT2
3
OUT3
4
OUT4
5
OUT5
6
OUT6
7
OUT7
8
All zero
Input
Delimiter
Output
Delimiter
Delimiter
ZERO 0 1
Force-zero is executed for OUT1.
OK
Input
Delimiter
Output
Delimiter