SECTION 3
Setting Measurement Conditions (Expert Menu)
Judgment Value Hysteresis Width
Set the hysteresis width for judgment upper/lower limit if the HIGH/PASS/LOW judgment does
not stabilize near the boundary.
Example: When HI is 100.000000 mm, LO is 80.000000 mm, and the judgment value hysteresis value is
0.050000 mm
Setting range
0.000000 to 999.999999
HI
䋨
100.00000
䋩
LO
䋨
80.00000
䋩
HIGH
OFF
ON
OFF
ON
OFF
ON
LOW
PASS
Hysteresis
䋨
0.050000
䋩
Hysteresis
䋨
0.050000
Reset at 80.050000mm
Output
Reset at 99.950000mm