![background image](http://html.mh-extra.com/html/omron/z550/z550_operation-manual_1633335164.webp)
162
S
E
CT
IO
N 4
Te
stin
g M
eas
ur
em
ent Pe
rfor
ma
nce
(Tes
t)
Z550
Operation Manual
SECTION 4
Other Functions
Using the TRIG Key to Specify the Start and End of Test
Measurements
1.
Perform the same operations as those described in 1. to 3. in the procedure for selecting [#
times] to display the test method selection menu.
p.160
A menu for selecting test methods appears.
2.
Set the [Test mode] to [Trg key].
3.
Select [End].
The focus returns to the test measurement screen.
4.
Display the output whose measurement value you want to check.
Up/Down key: Switches between outputs.
5.
Press the TRIG Key.
The testing starts and the screen display changes to
[RUN].
6.
Press the TRIG Key again in order to end the test
measurements.
OUT0
+000.2332mm [STOP]
MAX-MIN
No.
1
2
3
4
5
AVE
NUM
Scn 0
䂯
Tools
䂯
Test mode
䋺
Trg key
䂯
# of samples: [
䇭
䇭
100]
End
Press the TRIG Key again.
changes from [STOP] to [RUN].
During measurement, [RUN] flashes.
OUT0
MAX-MIN
No.
1
2
3
4
5
AVE
NUM
+000.2332mm [
RUN
]
Scn 0
䂯
Tools
䂯
operation.book Page 162 Monday, June 6, 2005 10:19 AM