SECTION 5
System Settings
Selecting How Height Coordinates are Handled
Select how the height coordinates are handled when the measurement range is changed.
Example: When the measurement range is changed from 60 mm to 30 mm
1.
Enter the System mode and select [Environment].
p.174
The setting screen appears.
2.
Change the setting of [Coordinate].
3.
Select [End].
The setting will be registered and the screen display returns to
1
.
Option
Meaning
Individual coordinates
(default setting)
A separate coordinate system is used for each measurement range.This option is
used under normal circumstances.
Common coordinates
One common coordinate system is used for all measurement ranges. (Because
images differ according to the measurement range, the measurement values may
change if the measurement range is changed.)
+30 mm
+24 mm
0 mm
-30 mm
Individual coordinates
+15 mm
+9 mm
0mm
-15mm
Common coordinates
+30 mm
+24 mm
+15 mm
0 mm
Scn 0
ٕ
System
ٕ
Comm
Profile
Environment
Download
Initialize
Version
Settings
Sensor
Scene
Monitor
1st scene
Coordinate
Dec. pts
Download
End
Scn 0
ٕ
System
ٕ
:
:
:
:
:
:
:
:
Expert
Mirror
Console
Image
Scn 0
Individual
6 digits
OFF
Common
Individual