Operation Manual
SECTION 3
Setting Measurement Conditions (Expert Menu)
A profile graph appears.
2.
Check that the workpiece is positioned correctly and
press the ENT Key.
The screen for setting height 1 appears.
3.
Specify the start and end lines.
Left/Right key: Moves the cursor.
(Press these keys together with the SHIFT Key to move
the cursor quickly.)
ENT Key: Selects the setting.
ESC Key: Cancels the setting.
The characteristic quantity setting screen appear.
4.
Select the characteristic quantity to be used as the
reference point.
Select which position (peak, bottom, or average) within the
specified area will be measured.
The cross cursor indicating the measurement point moves
according to the selection; therefore, select the position
while monitoring the image.
The screen for setting height 2 appears.
5.
Specify the characteristic quantity for the
measurement line and measurement point for the
second point using the same procedure as in
3.
to
4
.
Average
Peak
Bottom
Peak
Bottom
Average
Peak
Bottom