101
Z550
Operation Manual
S
E
CT
ION 3
Se
tting
Calib
ra
tion
SECTION 3
Setting Measurement Conditions (Expert Menu)
Example 2: In the case of [Edge position]
Edge position: 5 mm
Compensate for a measurement value of 4.95mm, so that it can be registered as
5 mm.
[Height] in Example 1 is used here to describe the setting method.
Make the necessary adjustments when setting a different item.
1.
Select [Meas set].
The selections for output appears.
2.
Move the cursor to the output for which the
calibration is to be set, and press the ENT Key.
Calibrate to 5 mm.
5mm
5mm
4.95mm
Measurement of point 1
Measurement of point 2
Move the workpiece by 5 mm
for the measurement.
Input the
position
of point 2.
5mm
Input the
position
of point 1.
0mm
Compensate for this error.
Measurement value
Workpiece Movement
Measure with the workpiece
at the reference position.
Scn 0
䂯䇭
Set
䂯
Image
Detect set
Meas set
Extra set
Meas mode
Output
Scn 0
䂯䇭
Set
䂯
OUT0:Height
OUT1:Step: 2 pts
OUT2:
OUT3:
OUT4:
OUT5:
OUT6:
OUT7: