3.3 FIRST SWITCH-ON
When power is applied the recorder initialises, and once this process is complete, the home page is
displayed. It is unlikely that this will contain any useful information because the input channels will not, as
yet, have been confi gured to suit the type of input signals being applied to them, as described in
section
4
.
Notes:
1 There is no on/off switch associated with the recorder.
2 Date, time and the message ‘ Power Up’ are printed on the chart each time power is applied to
the recorder, followed by a similar message giving ‘Confi g Revision’ and ‘Security Revision’.
3 A red line is drawn across the width of the chart at power up
The recorder has four security levels as follows:
Logged out Initially, no access to recorder confi guration is possible. Only Archive, Login/security and
the System ‘About’ functions can be accessed - via the root menu. Limited or full access
can be permitted from ‘Engineer’ level.
Operator*
No access to recorder confi guration is possible until access permissions have been set up.
Section 4.4.1
describes how limited or full access can be permitted by an operator with
‘Engineer’ level access.
Engineer*
Accessed initially, by entering ‘ 100’ as the password (section 3.3.1, below). Full access to all
recorder functions is available.
Section 4.4.1
describes how the Engineer level password
can be edited and an Operator password can be entered if required. The section also de-
scribes how access permission to some or all of the recorder functions can be granted, or
not, to individual user names and default security levels (except ‘Service’).
Service
Full access to all recorder functions and to areas of recorder memory for diagnostic pur-
poses. For use only by Service Engineers.
*Note: For units with the Auditor 21CFR11 Option enabled, the default Engineer and Operator